Dissolved Ozone Sensors
Robust and accurate inline and portable dissolved ozone analysers
Inline dissolved ozone sensor
The pureO3 dissolved ozone sensor provides rapid and accurate measurement across a wide range of ozone concentrations. It uses Intelligent Sensor Management (ISM) technology to provide predictive diagnostics for low-maintenance, in situ measurement.
Specifications – Dissolved Inline Ozone Sensor Pureo₃ | |
---|---|
Operating Rage – Short Term | 0 – 5000 ppb (µg/L) |
Operating Range – Continuous | 0 – 500 ppb (µg/L) |
Response time | 90% Response in 30 Seconds |
Sample Pressure (Metric) | 0.8 – 3.0 bar |
Sample Pressure (Imp) | 0 – 45 psig |
Sample Flow Rate with Housing | 200-500 mL/min |
Sample Flow Rate without Housing (Metric) | 0.15-1 m/s |
Sample Flow Rate without Housing (Imp) | 0.5-3 ft/s |
Connection Type | 1/4″ NPT(F) |
System Accuracy | ±1% of reading or 0.4 ppb, whichever is greater |
Sanitizable | Yes |
Portable Dissolved Ozone Sensor
DO3 is a rapid, simple to use device for checking the ozone concentration in water and is calibrated in mg/L (same as parts-per-million by volume in water). This small, battery-operated detector offers an inexpensive alternative to more costly inline monitors and requires no disposable reagents.
Specifications:
- Size: 3.13 x.1.56 x 1.36 inches (79.5 x 39.6 x 34.5 mm)
- Weight: < 2 ounces (49 grams)
- Range: 0.00 – 2.00 mg/L (ppm in water by weight), display indicates to 5.00 ppm, accuracy above 2 ppm is not implied
- Recommended water temp: 20 – 30 C (68 – 86 F)
- Operating Temp: 0 – 40 C (32-105 F)
- Accuracy: Within 20% when operated near room temperature.
- Response Time: 5 minutes to take final reading, but >90% value available on display after 3 minutes.
- Alarms: Buzzer
- Sampling Method: Gaseous Diffusion
- Interferences: Dissolved Chlorine
- Display: Digital LCD in increments of 0.01
- User Interface: Single button
- Tests: Self-Test on startup checks circuitry, alarms, battery, and operating temperature.
- Calibration: Recommended at least 1x per year, or whenever accuracy of reading is critical.
- Sensor: Transducer Technology T-Series electrochemical (3ET1PO3)
- Battery Information: Battery check on startup and during operation. User replaceable CR2450 coin battery.